This is a black testing report we host for developers to fill out when they are proposing updating the code on master.

Testing Report

Date Version Revision

Date:

Test Evaluator:

Item Description Result Comment
1. Software being evaluated. Include version, author, date,
2 Compile outcome
3 % of program space
4 % of dynamic memory
5 Zero’ing of airway pressure sensor
6 Zero’ing of flow measurement system
7 Ability to re-zero during later operation
8 Protection from accidentally turning Ventilator OFF
9 Display of Chosen Respiratory Rate
9B RR selectable
10 Display of Chosen I:E Ratio
10B I:E at least 1:1 1:2 1:3
11 Display of Chosen Peak Pressure Limit
12 Display of Chosen Flow Limits(upper lower or otherwise)
13 Display of airway pressure observed
13B Display of PEEP
14 Display of flow measured
15 Low pressure alarm
16 High Pressure alarm
17 Stop inspiration at high pressure alarm
18 Open expiratory valve at high pressure alarm
19 Alarm for low flow volume
20 Alarm for high flow volume
21-29 Reserved      
30 Alarm Temporary Silence
31 Inspiratory Pause Enable
32 Inspiratory Pause Disable
33 Assist Ventilation Enable
34 Assist Ventilation Disable
35 Trigger for Assist Ventilation
36 Delay from trigger to effective inspiratory assistance
40 Failure if button held down
41 Outcome of random simultaneous button presses
42 Outcome of pulling power
43 Outcome of pulling BMP280 clock line
44 Outcome of pulling BMP280 data line
45 Outcome of pulling BMP280 Vin line
46 Function of ventilator with missing BMP280glue (cold start)
47 Technique to re-enable BMP280
48 Outcome of attempting to re-enable BMP
49-50 Reserved      
50 Measured RR @ setting 10
51 Measured RR @ setting 30
52 Measure of idle time of processor
53 Active PEEP
54 Assist Vent